Sensor and Flow Train
· Sensor : Polarographic, Gold/Silver, PTFE membrane
· Range :
- Chlorine : 0.05 to 20 ppm
- pH : 0 to 14 pH
· Temperature : 32° to 122°F (0° to 50°C)
· Flow :
- 38 L/hr to 300 L/hr
- (10 gal/hr to 80 gal/hr)
· Wetted Materials : PVC, PP, PVDF, PTFE, Glass, 316 SS
· Process Connections : Input 1/4" barb fitting, drain 3/4" barb fitting
· Response Time : T90 in approximately 2 minutes
· Electrolyte Life : Up to 12 months
Electronics
· pH Compensation :
- Chlorine : 0.05 to 20 ppm
- pH : 0 to 14 pH
- Temperature : 32° to 212°F (0° to 100°C)
· Measurements : pH 5-10
· Display : 2.5" x 1.75" back-lit LCD, 4 lines for Text and Graphical
· Enclosure : NEMA 4X, 5.7"L x 5.7"W x 7"D
· Output :
- 4-20 mA, 800 ohm at 24 VDC
- Optional PID output and additional 4-20 mA outputs
· Input Power :
- 110 / 220 VAC at 50/60 Hz
- Optional 24 VDC (12 to 50 VDC) at 0.25A
· Alarm Relay Ratings :
- (2) SPDT 230 VAC / 5A or 30 VDC / 5A resistive max
- Optionally up to (8) relays
· 유리염소 센서
유리염소는 차아염소산(pH 5에서 ~ 100%)과 차아염소산염 이온(pH 10에서 ~ 100 %)의 pH 의존 비율로 용액에 존재합니다.
유리염소 센서는 유리염소의 차아염소산 성분만 측정하고, 분석기는 측정된 pH 또는 사용자 정의 고정 값을 사용하여 균형을 계산합니다.
pH 센서를 사용하면 pH 6에서 pH 9.5 사이의 샘플을 정확하게 보상할 수 있으며 용액의 pH를 제어하기 위해 값비싼 샘플 컨디셔닝 시스템이 필요하지 않습니다.
LXT-330은 사용자 정의 범위를 사용하여 두 파라미터 중 하나를 그래픽으로 표시하여 쉬운 트렌드 분석을 가능하게 합니다.
· 전류계염소 센서
전류계염소 센서는 유량에 민감하며 센서에 의해 필요한 최소 유량은 0.5 ft / sec이며, 이 값보다 높으면 출력은 유량에 거의 독립적입니다.
"일정한 헤드" 유량 제어장치(CFD)는 광범위한 유입 샘플 유량에서 센서에 의해 최적의 유량을 유지합니다.
CFD에 필요한 최소 유량은 10 gal / hr이고, 최대 유량은 샘플이 대기압에서 배출될 때 80 gal / hr입니다.